40
Connect ECG Via Analog Input AUX IN 2
An ECG signal may be imported from the HRS and displayed in the
Live
window on the
QUANTIEN™ measurement system.
Connection:
Connect an AUX IN (ECG) cable between an analog ECG output port on the HRS and AUX IN 2
(Product Interface and Symbols (page 3)) on the back of the QUANTIEN main unit. The cable
used must be compatible with the specific HRS.
Setup:
Setup ECG in:
Select Room, New Room / Edit Room, AUX IN 2
ECG Configuration:
Scale
button. Select ECG scale.
Select
Display Trace
checkbox to display ECG signal.
Select
Baseline Filter
checkbox to enable high pass filter for baseline drift removal.
Connect PressureWire™ Output to Recording System
WARNING: The PressureWire™ OUT output provides isolation according to
defibrillation proof type CF (see figure Isolation Diagram (page 78)). External
equipment which is connected to this output must provide isolation against
leakage current. The user is responsible for compliance with the requirements of
standard IEC 60601-1-1, Safety requirements for medical electrical systems,
alternatively IEC 60601-1 ed.3§16, ME SYSTEMS.
Connection:
To display Pd pressure on a HRS connect a monitor cable from PressureWire OUT (Product
Interface and Symbols (page 3)) to a free input port on the HRS.
Configuration:
To ensure equal pressure readings on the HRS and the QUANTIEN™ measurement system, the
associated input port must be zeroed. Zero the associated input port after PressureWire™
guidewire has been zeroed on the QUANTIEN measurement system and before inserting
PressureWire guidewire into the patient, or by using the static 0 mmHg reference output setting in
(Setup)
Menu,
System, Other, PW Reference out
.
Settings:
Pd
– Distal pressure from wireless or cable connected PressureWire guidewire
0 mmHg
– static 0 mmHg reference pressure
100 mmHg
– static 100 mmHg reference pressure
Configure Analog Output Ports AUX OUT 1 & 2
Connection:
AUX OUT 1 and AUX OUT 2 (Product Interface and Symbols (page 3)) are analog output ports
enabling connection of external measurement devices using an AUX out cable.
Configuration:
Select Room, New Room / Edit Room, AUX OUT 1 or 2
Settings:
Pd
– Distal pressure from wireless or cable connected PressureWire™ guidewire
Pa
– Proximal pressure from AO transducer
Reference voltage levels
: 0V; 1.0V; 1.5V
State: Released
Date: 2019.03.06 22:30 GMT