39
-
Press
Zero and Calibrate
– the Dialogue box in the figure below opens.
-
Make sure that the AO pressure channel on the cath lab HRS is zeroed.
-
Press the
Zero
button in the left column. The AO signal is now zeroed.
Figure 21. Zero and Calibrate AO
1.
Zero Status
2.
Zero Offset
3.
Active menu:
Zero and Calibrate
4.
Nominal Sensitivity
5.
Calibration Status
6.
Calibration Value
7.
Reference Level
Default Calibration:
If the sensitivity of the analog output port on the HRS is fixed at 100 mmHg/V the nominal
sensitivity can be used. Check the checkbox
Use Nominal Sensitivity
.
Custom Calibration:
For all other system setup or if there is uncertainty regarding the nominal sensitivity, an
arbitrary reference level from the HRS may be used to calibrate the AO pressure input port.
Enter the reference level used from the HRS in the field
Reference Level (mmHg)
and
press the
Calibrate
button. The system is now calibrated and ready to receive AO pressure
from the HRS.
The calibration factor and date are saved for future reference.
State: Released
Date: 2019.03.06 22:30 GMT