INTRODUCTION
T
EMPERATURE
R
EFERENCE
S
ENSOR
D
EVICE
Must be purchased separately
A reference sensor device is required for calibrating the incubator temperature display.
Reference devices must meet the following standards:
•
Accurate to at least 0.1°C
The device should be regularly calibrated, preferably by a third party.
Temperature Probe
Use a digital device with a wire thermocouple probe that can be introduced into the incubation chamber
through the unit access port. Select a thermocouple suitable for the application temperature you will be
calibrating at.
Why a Probe?
Reference readings taken outside the chamber using wire temperature probes avoid chamber door
openings. Openings disrupt the chamber temperature. Each disruption requires
a minimum 1-hour wait
to allow the atmosphere to re-stabilize before continuing.
No Alcohol or Mercury Thermometers
Alcohol thermometers do not have sufficient accuracy to conducti accurate temperature calibrations.
Never place a mercury thermometer in the incubation chamber!
Always use thermocouple probes.
Temperature Reference