OPERATION
Note:
Make sure the day – night auto cycle is not running prior setting the manual mode
temperature set points. A red graph symbol will be illuminated on the right side of the controller
homepage if the cycle is active.
S
ET THE
M
ANUAL
N
IGHT
M
ODE
T
EMPERATURE
S
ET
P
OINT
Optional
: The indefinite constant temperature manual night mode is set to 20°C at the factory.
Change this set point if you will be running the manual night mode at another application
temperature.
Perform steps 1 – 8 whenever required to put the unit in the manual night mode.
Continued on next page
Put the unit in manual night mode
1.
Verify the unit is running the manual day mode before attempting
to place it in the night mode. Indicators 2, 3, and 5 will be
illuminated in the manual day mode.
•
If the auto day – night cycle is running, turn it off by
pressing the EZ1 button.
•
If the unit is already in the manual night mode, skip to
step 9 to set the night set point. When in manual night
mode the small “6” indicator near the bottom of the
screen will be illuminated.
Manual day mode active,
diurnal cycle off.
2.
Navigate to the operations menu.
a.
Press and hold
both the Up and Down Arrow
buttons
for approximately 5 seconds.
•
The display will change to show the Operation Options
menu page.
Operations Menu Page
3.
Scroll through the Operations menu to the Parameters option.
a.
Press the
Up Arrow
once.
Parameters Option
Cycle Active