USER MAINTENANCE
Continued on next page
Temperature Display Calibration – Day Temperature
1.
Once the day temperature has stabilized, compare the reference device
temperature reading and the reading of the chamber controller.
a.
If the readings are the same, or a difference between the two falls
within the acceptable range of your protocol, the display is
calibrated for day temperature.
Advance to the Calibrating the
Night Temperature steps starting on page 52.
- OR -
b.
If a difference falls outside your protocol range, advance to the
next step.
Reference Device
2.
A display calibration offset must be entered to match the current display
value to the reference device reading.
Reference Device
3.
Determine the offset value and write down for use in Step 5.
•
The offset value is the difference between the reference device
reading and the reading of the chamber controller temperature
display.
Reference Sensor
Reading
Controller
Display
Offset
Value
23.0°C
22.0°C
1.0
21.2°C
22.0°C
-0.8
Reference Device
Offset Value: -0.8
4.
Navigate to the Operations menu page.
a.
Push and hold
both
the Up and Down Arrow Button
simultaneously for approximately 5 seconds.
Note:
Holding down both arrows for 10 seconds will open a Set menu
for setting memory parameters. If this happens, press the Reset
button to return to the Home page, the begin this step again.
Operations Menu