USER MAINTENANCE
Procedure continued on next page
Night Temperature Calibration (Continued)
7.
Scroll to the Calibration Offset option
a.
Press the green Advance button 3 times until “i.CA” appears in
the green middle display line.
Current Offset 0
8.
Enter the current calibration offset.
a.
Use the Up and Down arrow keys to enter the offset in the red
numerical display line.
Current
Offset, 1
9.
Once the offset is entered, push the Reset Button (∞) twice.
•
The first push exits the Operation Options menu and saves the
current offset in the controller short-term memory.
•
The second push returns the display to the home page, now
showing current display value corrected with the offset to
match the reference device reading.
•
The chamber will now heat or chill to match the set point with
the corrected display value.
Chamber cooling to
achieve the set point
with the corrected
display value
10.
Once the chamber has achieved the set point temperature, wait at least 1
hour for the chamber to stabilize.
a.
Failure to wait until the cultivation chamber is fully stabilized
will result in an inaccurate reading.
x3
x2