INSTALLATION
R
EQUIRED
A
MBIENT
C
ONDITIONS
The SRI chambers are intended for use indoors at room temperatures between
15
°
C and 30
°
C
(59
°
F and 86
°
F)
, at no greater than an ambient
80% Relative Humidity
(at 25
°
C / 77
°
F). Allow a
minimum of
4 inches (100 mm)
between the unit and walls or partitions, and
2.5 inches (60 mm)
of
clearance above the top of the chamber for unobstructed airflow.
Operating the unit outside these conditions may adversely affect its temperature range and stability.
E
NVIRONMENTAL
D
ISRUPTION
S
OURCES
When selecting a location to install the chamber, consider all environmental conditions that can
affect the unit temperature performance. For example:
•
Proximity to ovens, autoclaves, and any device that produces significant radiant heat
•
Heating and cooling ducts, or other sources of fast-moving air currents
•
High-traffic areas
•
Direct sunlight