OPERATION
P
ROGRAM
C
LOCK
T
IMES AND
C
YCLE
T
EMPERATURE
S
ET
P
OINTS
A
quick reference chart
for this procedure is available at its end on page 37.
Steps 1 – 4 of this procedure set the controller clock to your local time. The clock runs on a 24-hour
format. See the table below for examples of 24-hour and am – pm equivalencies.
Steps 5 – 17 set the start times and temperature set points for the day and night phases of the auto
illumination and temperature cycle.
Skipping steps
: If do not wish to change a time setting or temperature set point, do not use the
arrow keys to change the setting. Instead, press the Advance key to scroll through a step without
adjusting it.
Go back a step
: To go back to the start of this procedure, press the Reset button once. Pressing the
button twice at any point will return to the Home page.
Procedure continued on next page
12-Hour Clock
24-Hour Controller
Clock
12-Hour Clock
24-Hour Controller
Clock
8:00AM
08:00
2:00PM
14:00
10:00AM
10:00
6:00PM
18:00
12:00PM
12:00
8:00PM
20:00
1:00PM
13:00
12:00AM
00:00
Setting the chamber clock to local time
1.
Starting on the home page, scroll to the current clock time hour
page.
a.
Press the
Advance button
once.
•
“hoU1” is the clock hour memory slot (set to 1 pm or 13:00
in this example).
Note:
The controller display example to the right will show an adjustment
from 1:34 pm to 9:45 am.
Initial Clock Time Hour
2.
Set the clock time hour.
a.
Use the
Arrow keys
to adjust to the hour to the current
local hour (9 am in this example).
Adjusted Clock Time Hour