INSTALLATION
D
EIONIZED AND
D
ISTILLED
W
ATER
Do not use deionized water
to clean the chamber. Use of deionized water may corrode metal
surfaces and voids the warranty. Sheldon Manufacturing recommends the use of distilled water in
the resistance range of 50K Ohm/cm to 1M Ohm/cm, or a conductivity range of 20.0 uS/cm to 1.0
uS/cm, for cleaning applications.
I
NSTALLATION
C
LEANING AND
D
ISINFECTION
If required by your laboratory protocol, clean and disinfect the cultivation chamber and shelving
components prior to installation. Cleaning and disinfecting reduce the risk of contamination. The chamber
and shelving were cleaned and disinfected at the factory, however, Sheldon Manufacturing cannot
guarantee that the chamber was not exposed to contaminants during shipping.
Remove all protective wrappings from shelving components prior to cleaning.
Please see the
Cleaning and Disinfection procedure
on page 41 of the User Maintenance section
for information on how to clean and disinfect without damaging the chamber and its components.