USER MAINTENANCE
Figure 16: Summary chart for calibrating night temperature
Prior to calibrating, verify that
6
is illuminated, indicating the unit is in the
manual night mode
.
19.0
19.0
Push
and
hold
both
buttons:
5
seconds
Push 3
times
i.CA
0.0
Enter the
calibration
offset
using the
arrow
buttons
i.CA
1.0
Push Reset
twice
19.0
19.0
Return to
Home Page
A1
oPEr
Saving the current calibration value (night temp) in the long-term memory
A1
SET
19.0
19.0
Push 3
Times
LbL
g
SET
Push 15
times
Or until of fset
memory slot s
appears
nonE
Usr.S
Usr.S
SET 1
Push once
Push Reset
twice
19.0
19.0
Return to
Home Page
Selecting the Set 1 memory slot and pressing RESET saves the current calibration offset (night) in the in the long-term memory
Operations Menu
RES E T
Current Calibration Value
Home
Page
Home Page
Home Page
Push
and
hold
both
buttons:
10
seconds
Memory Parameters Menu
Global Parameters List
Offset Memory Slots
Night Cal. Memory Slot
RES E T
Home Page
End of procedure