USER MAINTENANCE
Procedure continued on next page
Saving the Night Calibration (Continued)
13.
Navigate to the Settings Menu page (SEt).
a.
Push and hold
both
the Up and Down Arrow Button
simultaneously for 10 seconds.
•
The display will advance to the Operations page after
approximately 5 seconds, then to the Settings Menu page
after another five seconds.
Settings Menu Page
14.
Advance to the Global Settings parameter list.
a.
Press the Up arrow button 3 times.
Global Settings List
15.
Scroll to the Memory parameter.
a.
Press the
green Advance button
15 times.
Temperature Slot
Parameters
16.
Select the night temperature calibration “Set 1” memory slot.
a.
Press the
Up Arrow
button
1 time.
Note:
Select the None memory slot if you do not wish to save the current
calibration value.
Night Cal. Memory Slot
x3