53 |
P a g e
USER MAINTENANCE
Procedure continued on next page
Night Temperature Calibration (Continued)
4.
A display calibration offset must be entered to match the
current display value to the reference device reading.
Reference Device
5.
Determine the offset value and record it for use in three steps.
•
The offset value is the difference between the reference
device reading and the reading of the chamber controller
temperature display.
•
The offset value will be used to adjust the controller
reading to match the reference device.
Reference Sensor
Reading
Controller
Display
Offset
Value
20.0°C
19.0°C
1.0
18.0°C
19.0°C
-1.0
Reference Device
Chamber Display
Matching Offset Value: 1.0
6.
Navigate to the Operations menu page.
a.
Push and hold
both
the Up and Down Arrow
buttons simultaneously for approximately 5
seconds until the oPEr menu appears.
Note:
Holding down both arrows for 10 seconds will open a Set menu
for setting memory parameters. If this happens, press the Reset button
to return to the Home page, then press both Arrow buttons again for 5
seconds.
Operations Menu