26 |
P a g e
OPERATION
Note
: Make sure the day – night auto cycle is not running prior to setting either of the manual mode
temperature set points. A red graph symbol will be illuminated on the right side of the controller
homepage if the cycle is active. See the instructions below for terminating the cycle.
S
ET THE
M
ANUAL
D
AY
M
ODE
T
EMPERATURE
S
ET
P
OINT
Optional
: The indefinite constant temperature manual day mode is set to 20°C at the factory.
Change this set point if you will be running the manual day mode at another application temperature.
The chamber comes from the factory set to start in the manual night mode. Perform steps 1 – 8
whenever required to put the unit in the manual day mode.
Continued on next page
Put the unit in manual day mode
1.
Verify the unit is running the manual night mode before attempting to
put it in the day mode. Indicator 6 near the bottom of the screen will be
illuminated in the manual night mode.
•
If the auto day – night cycle is running, turn it off by
pressing the EZ1 button.
•
If the unit is already in the manual day mode, skip to step 9
to set the day set point. When in manual day mode the
small 2, 3, and 5 indicators near the bottom of the screen
will be illuminated.
Night Mode Active, diurnal
cycle off.
2.
Navigate the operations menu.
a.
Press and hold
both the Up and Down Arrow
buttons for
approximately 5 seconds.
•
The display will change to show the Operation Options
menu page.
Operations Menu Page
3.
Scroll through the Operations menu to the Parameters option.
a.
Press the
Up Arrow
once.
Parameters Option
Cycle Active