USER MAINTENANCE
Unit Performance Specifications
What is the unit designed to do?
Find and note the following unit designation and performance information in the user manual.
Unit Specification
Data Location
Record Data Here
Model Type:
The unit data plate lists the model
type. Verify you are attempting to
diagnose an SRI21D.
What is the operating
temperature range of the unit?
The Temperature block in the Unit
Specifications chapter, page 68.
What is the required ambient
temperature range for the unit?
See below (under this table).
What is the minimum time
required for your unit to come up
to temperature and stabilize?
Allow a minimum of 4 hours of
continual operation for the unit to
achieve a stable temperature.
Standard ambient environment temperature requirements:
•
A room temperature of 15°C - 30°C (59°F - 86°F)
Standard Ventilation Spacing Requirements
•
A minimum of
4 inches (100 mm)
clearance between the sides of the cultivation chamber and any
walls or partitions. A minimum of
25 inches (50 mm)
between the top of the unit and overhead
cover.
This page may be copied for institutional use