USER MAINTENANCE
A
mbient?
What is the room temperature?
•
For best results, measure the temperature in the same section of the room where the unit is
located. Do not place your thermometer on the unit.
P
ilot Lights?
1) Is the heating active indicator on the control panel flashing or otherwise illuminating.
During normal operations, the “1” indicator should flicker on and off frequently as the heating
element competes with the refrigeration compressor to achieve a stable temperature within ±0.5°C.
The compressor should run continually when the unit is on.
Refrigeration
: Please listen to hear if the refrigeration compressor is running or not. Note the
absence of compressor noises on the SDRAP form. The operation of the compressor in a noisy
environment can also be detected from feeling for vibrations on the lower back of the unit.
Cooling issues can also be generated by a failure of the blower fan in the cultivation chamber.
2)
Is the Over Temperature Limit Activated indicator illuminating? Answer yes if the light is on
continually or if it is pulsing.
“1” indicates the
controller is calling for
power to the element(s)
Figure 19: Heating Indicator
OVER TEMPERATURE ACTIVATED
Room Air Temperature