USER MAINTENANCE
Figure 15: Summary chart for calibrating day temperature
Prior to calibrating, verify that
2
,
3
, and
5
are illuminated, indicating the unit is in the
manual day mode
.
22.0
22.0
Push
and
hold
both
buttons:
5
seconds
Push 3
times
i.CA
0.0
Enter the
calibration
offset
using the
arrow
buttons
i.CA
-0.8
Push Reset
twice
22.0
22.0
Return to
Home Page
A1
oPEr
Saving the current calibration value (day temp) in the long-term memory
A1
SET
20.0
20.0
Push 3
Times
LbL
g
SET
Push 15
times
Or until of fset
memory slot s
appears
nonE
Usr.S
Usr.S
SET 2
Push once
Push Reset
twice
22.0
22.0
Return to
Home Page
Selecting the Set 2 memory slot and pressing RESET saves the current calibration offset (day temp) in the in the long-term memory
Operations Menu
RES E T
Current Calibration Value
Home
Page
Home Page
Home Page
Push
and
hold
both
buttons:
10
seconds
Memory Parameters Menu
Global Parameters List
Offset Memory Slots
Day Cal. Memory Slot
RES E T
Home Page
Saving the Night Calibration (Continued)
17.
Save the current calibration value in the memory slot and return to the
Home Page by pressing the
Reset Button
(∞) twice.
o
The first Reset Button press saves the current calibration value in
the presently selected calibration long-term memory slot (Set 1).
o
The second button press scrolls the display back to the Home
Page.
•
The chamber is now calibrated for temperature. Return the unit to
normal operations.
Home Page
X2
Содержание SRI21D
Страница 1: ...DIURNAL PLANT GROWTH CHAMBER 115 Voltage Installation Operation Manual SRI21D ...
Страница 8: ...8 P a g e INTRODUCTION ...
Страница 12: ...12 P a g e RECEIVING YOUR UNIT ...
Страница 20: ...20 P a g e GRAPHIC SYMBOLS ...