USER MAINTENANCE
Procedure continued on next page
Night Temperature Calibration (Continued)
11.
Compare the reference device reading with the chamber
temperature display again.
a.
If the reference device and the chamber temperature display
readings are the same or the difference falls within the range of
your protocol,
advance to Saving the Night Calibration on
page 56
.
- Or -
b.
If the readings fail to match or fall outside your protocol range,
advance to the next step.
Reference Device
12.
Repeat steps 3 – 11 up to two more times for the night phase if the
unit is still not calibrated.
•
Three calibration attempts may be required to successfully
calibrate units that are more than ±2°C out of calibration.
•
If the readings of the current temperature display and the
reference device are still not close enough to be acceptable by
your laboratory protocol after three calibration attempts,
contact your distributor or technical support for assistance.
Reference Device
Saving the Night Calibration
Once the display reading matches the reference device, the current offset
must be saved in the night temperature calibration memory slot or it will be
lost the next time the chamber is turned off.