4 Customizing Your System
Basic User Manual 41
Item
Description
Remote HostName
Enter the name of the DICOM worklist server.
IP Address
Set the IP address of the DICOM worklist server.
Click
Ping
to test the connection.
DICOM AE Title
Set the Application Entity (AE) title of the DICOM worklist server.
Port Number
Set the port number of the DICOM worklist server.
Connect Timeout(sec)
Set the time limit for the network connection timeout.
Repeat Count
Set the number of repeated attempts for connecting to the DICOM worklist server.
Dimse Timeout(sec)
Set the timeout time for sending the Dimse command.
Acse Timeout(sec)
Set the timeout time for sending the Acse command.
Max. Result
Set the maximum number of patient records that can be sent to the ultrasound
system.
Click
Echo
to verify the connection between the ultrasound system and the DICOM worklist after you input the
settings for the
Remote HostName
,
IP address
,
DICOM AE Title
,
Port Number
. A summary displays if the
connection is verified.
4�7�4
MPPS Settings
You can connect the ultrasound system to the MPPS server and adjust the relevant settings on the MPPS tab.
System Setting
Store
C-Store
Worklist
MPPS
Delete
Add
Service List
MPPS
DICOM MPPS
Remote HostName
MPPS
Port Number
109
Connect Timeout(sec)
10
Repeat Count
2
Dimse Timeout(sec)
10
Acse Timeout(sec) 30
Ping
Echo
IP Address
192.168.254.104
DICOM AE Title
AE
Peripheral
General
Comment
Measure
DICOM
Load Default
About
Exit
Save&Exit
Bodymark
Report
Figure 4-19
DICOM Settings Menu - MPPS Tab
Содержание EVUS 8
Страница 1: ...C d Rev 02 77000001436 EVUS 8 OWNER S MANUAL English...
Страница 10: ...This page is intentionally left blank...
Страница 18: ...This page is intentionally left blank...
Страница 62: ...This page is intentionally left blank...
Страница 88: ...This page is intentionally left blank...
Страница 92: ...This page is intentionally left blank...
Страница 112: ...This page is intentionally left blank...
Страница 122: ...This page is intentionally left blank...
Страница 149: ...139 Appendix E Acoustic Output Data Please refer to Section 4 9 2 Acoustic Output...
Страница 150: ...NUM REG ANVISA 10069210070 www saevo com br...