13 Probes and Biopsy
Basic User Manual 119
2. Perform the medium-level disinfection to the probe cable as shown in Table 13-4.
3. Rinse the surfaces of the probe cable with fresh running water to completely remove the residual disinfectant.
4. Dry the probe cable with a lint-free soft dry cloth.
Table 13-4
Recommended Method for Medium-level Disinfection
Disinfectant
Manufacturer
Active Ingredients
Level of Active
Ingredients
Contact
Type
Contact Period
T-spray II
Pharm. Inc.
Quat. Ammonia
/
S p r a y /
Wipe
<10 minutes
T-spray
Pharm. Inc.
Quat. Ammonia
/
S p r a y /
Wipe
<10 minutes
■
Sterilization
If the probe cable comes into contact with any blood or body fluid, you should sterilize it. For details, refer to
Section 13.1.5 Sterilizing the Probe.
13�1�7 Storage and Transportation
You should operate, store and transport the probe in accordance with Appendix A Specifications.
■
To store the probe
Follow the following items to store the probe.
●
Store the probe in a probe holder or specific package when not in use.
●
Ensure the probe holder is clean before storing it.
●
To avoid contaminating the package, disinfect or sterilize the probe before storage or store it in a disposable
probe package.
●
Do not expose it to direct sunlight or sudden environmental temperature changes.
●
Store the probe alone to avoid any impact on the probe head.
●
Carefully wind the probe cable.
●
Dry the probe before storing it.
■
To transport the probe
Perform the following steps to transport the probe.
1. Clean and disinfect the probe before transportation.
2. Place the probe fully into the carrying case and carefully wind the cable.
3. Close the cover of the carrying case.
4. Pack the carrying case with sponge and place it in a carton.
13�2
Biopsy
You should properly assemble and verify the biopsy bracket before performing a biopsy.
Содержание EVUS 8
Страница 1: ...C d Rev 02 77000001436 EVUS 8 OWNER S MANUAL English...
Страница 10: ...This page is intentionally left blank...
Страница 18: ...This page is intentionally left blank...
Страница 62: ...This page is intentionally left blank...
Страница 88: ...This page is intentionally left blank...
Страница 92: ...This page is intentionally left blank...
Страница 112: ...This page is intentionally left blank...
Страница 122: ...This page is intentionally left blank...
Страница 149: ...139 Appendix E Acoustic Output Data Please refer to Section 4 9 2 Acoustic Output...
Страница 150: ...NUM REG ANVISA 10069210070 www saevo com br...