3 Preparing the System
20
Basic User Manual
3�3
Powering On/Off the System
●
Do not power off the system during a system upgrade or a data transmission.
●
An extra 30-60 seconds are needed for powering on the system if it is powered off in an abnormal way
previously.
●
If the system is configured with the μScan feature, do not connect the USB drive to the system before system
startup.
●
Wait approximately 20 seconds before powering on the ultrasound system again after powering it off.
Otherwise, a failure may result.
●
To maintain the safety and functionality of the ultrasound system, regular maintenance must be performed as
described in Chapter 14 System Maintenance.
●
Ensure the system works in accordance with requirements in
.
■
To power on the system
1. Connect to the adapter to system and the mains supply.
2. Press the button on the left side of the system to power on the system.
All operating indicators illuminate, the initial screen and the application mode screen appear after the system
completes the initialization.
■
To power off the system
Press the button on the left side of the system to power off the system.
■
System standby
If the ultrasound system is idle for a set period, the system enters the standby mode automatically.
3�4
Adjusting the System
Grip the left and right sides of the screen and swivel the screen forwards or backwards.
3�5
Connecting the Probe
Disconnect the probe from the system after freezing an image or powering the system off. Otherwise, the probe of
the system could be damaged.
■
To connect the probe to the system
1. Switch the probe lock above the probe port to the
position, and then insert the probe connector into the
probe port firmly.
2. Switch the probe lock to the position to lock it securely.
If the probe cable is too long, hang it on the probe cable hanger.
■
To connect the probe to the transducer extender
1. Switch the probe lock on the system to the
position, and then insert the probe connector of the transducer
extender into the probe port firmly.
2. Switch the probe lock to the position to lock the extender securely.
3. Switch the probe lock on the transducer extender to the
position, and then insert the probe connector of into
the probe port of the extender firmly.
4. Switch the probe lock to the position to lock the probe securely.
Содержание EVUS 8
Страница 1: ...C d Rev 02 77000001436 EVUS 8 OWNER S MANUAL English...
Страница 10: ...This page is intentionally left blank...
Страница 18: ...This page is intentionally left blank...
Страница 62: ...This page is intentionally left blank...
Страница 88: ...This page is intentionally left blank...
Страница 92: ...This page is intentionally left blank...
Страница 112: ...This page is intentionally left blank...
Страница 122: ...This page is intentionally left blank...
Страница 149: ...139 Appendix E Acoustic Output Data Please refer to Section 4 9 2 Acoustic Output...
Страница 150: ...NUM REG ANVISA 10069210070 www saevo com br...