17
3
Preparing the System
System preparation is necessary before using the ultrasound system. The preparation includes but is not limited to positioning
and adjusting the system, connecting the probe and the peripheral devices.
3�1
Positioning the System
Leave at least 20cm at the back and both sides of the ultrasound system for ventilation. Otherwise, the temperature
rise could cause failures.
Place this system on a level surface.
You can place the system on the trolley (sold separately) provided by manufacturer. For details on positioning the
system, please refer to the relevant user manual.
3�2
Powering the System
Two methods can be used to supply power for the system: mains supply and built-in rechargeable lithium battery.
When the adapter is connected to the mains supply, the battery is charged until it is full.
3�2�1
Using the Mains Supply
!
●
The accompanying AC adapter can only be used for power supply or battery charge of the system. Other usages
may damage the adapter or even cause an electric shock.
●
Ensure that the adapter is not damaged before use. If damage exists, stop using it immediately and contact the
manufacturer or the local distributor.
●
Use only the adapter provided by the manufacturer.
●
Do not hang the adapter in the air.
●
The adapter should be placed in a well ventilated area during use. Do not cover the AC adapter with any other
objects.
■
To connect the adapter
1. Connect the grounding terminal to the protective ground using the power grounding cable.
2.
Connect one end of the adapter to the system and connect the other end to the mains supply firmly.
After use, disconnect the cable from the mains supply and then disconnect the grounding cable.
3�2�2
Using the Battery
!
●
Do not disassemble or alter the battery. Otherwise, there is a danger of explosion.
●
Replace the battery with the type recommended by the manufacturer.
●
Do not short-circuit the battery by directly connecting the system with metal objects.
Содержание EVUS 8
Страница 1: ...C d Rev 02 77000001436 EVUS 8 OWNER S MANUAL English...
Страница 10: ...This page is intentionally left blank...
Страница 18: ...This page is intentionally left blank...
Страница 62: ...This page is intentionally left blank...
Страница 88: ...This page is intentionally left blank...
Страница 92: ...This page is intentionally left blank...
Страница 112: ...This page is intentionally left blank...
Страница 122: ...This page is intentionally left blank...
Страница 149: ...139 Appendix E Acoustic Output Data Please refer to Section 4 9 2 Acoustic Output...
Страница 150: ...NUM REG ANVISA 10069210070 www saevo com br...