10 Working with Images
100
Basic User Manual
10�6�1
Basic Procedures of ECG Operation
Perform the following steps.
1.
Power off the ultrasound system and connect the ECG cable to the specific port of the ultrasound system.
2. Power on the ultrasound system and attach the ECG electrodes on the patient’s body (as shown in the following
figure).
Yellow
Green
Black
Red
R
L
N
F
RA
LA
RL
LL
Green
Red
White
Black
IEC Standard
AHA Standard
I
I
II
II
III
III
Figure 10-6
Electrode Placement
3. Make the relevant settings of the ECG module. For details, refer to Section 10.6.2 Optimizing Parameters.
4. Freeze the images and review them. For details, refer to Section 10.6.3 Reviewing ECG.
5. Exit the ECG mode, and remove ECG electrodes from the patient.
10�6�2
Optimizing Parameters
Select the
Menu
key and then select
ECG
on the pop-up menu. Rotate the
Menu
key or select the ← or → key to
display the submenu of the ECG module.
Item
Description
ECG
Enable or disable the ECG function.
ECG Gain
Set the amplitude of the ECG waveform.
ECG Position
Set the vertical position of the ECG waveform.
ECG Invert
Enable or disable the waveform inversion function.
R-Trigger
Enable or disable the R-trigger function.
When enabled, the
R-Trigger
menu pops up, including several items to be chosen.
NOTE:
In order to enable the R-trigger function, you should disable the compounding
imaging function first.
Trigger Delay
Delay of trigger after the 2D refresh.
Frame Count
Set the number of the image transmitted in one second.
Frame Int.
Set the refresh interval of 2D image.
Содержание EVUS 8
Страница 1: ...C d Rev 02 77000001436 EVUS 8 OWNER S MANUAL English...
Страница 10: ...This page is intentionally left blank...
Страница 18: ...This page is intentionally left blank...
Страница 62: ...This page is intentionally left blank...
Страница 88: ...This page is intentionally left blank...
Страница 92: ...This page is intentionally left blank...
Страница 112: ...This page is intentionally left blank...
Страница 122: ...This page is intentionally left blank...
Страница 149: ...139 Appendix E Acoustic Output Data Please refer to Section 4 9 2 Acoustic Output...
Страница 150: ...NUM REG ANVISA 10069210070 www saevo com br...