6 Acquiring Images
68
Basic User Manual
To set the display format:
1. Select the
Menu
key in the real time scan. A menu pops up on the screen.
2. Select
Dis Format
by using the trackball or the ↑ or ↓ key.
3. Rotate the
Menu
key or select the ← or → key to adjust the setting.
■
Power
Power is used to select the amount of ultrasound acoustic power produced by the probe. The adjustment range of
the power is 1-100%, and ±10% can be adjusted each time. The real time value of the power is displayed in the
imaging information area on the basic screen.
To adjust the power:
1. Select
Power%
by pressing the corresponding function key.
2. Rotate it to adjust the value.
NOTE:
Expose the patient to the lowest practical transmit power level for the shortest possible time to achieve a
satisfactory diagnosis.
■
Video Invert
Video invert is used to invert the M trace display related to brightness.
To enable or disable the video invert:
1. Select
Video Invert
by pressing the corresponding function key.
2. Rotate it to enable or disable the feature.
6�5
Acquiring Spectral Doppler Images
Spectral Doppler imaging is intended to provide measurement data concerning the velocity, the direction, and the
category of the arterial or vein flow. It contributes a more accurate qualitative analysis than the color flow imaging.
Spectral Doppler includes Pulsed Wave Doppler (PW) mode and Continuous Wave Doppler (CW) mode.
6�5�1
PW Mode
Pulsed Wave Doppler (PW) is a Doppler mode that measures velocity in a PW sample volume and displays
that information in a spectral trace with audio output. PW includes conventional PW and High Pulse Repetition
Frequency (HPRF). PW Doppler can be combined with the B mode for rapidly selecting the anatomical site for PW
Doppler examination. The site where PW Doppler data is derived appears graphically on the B-Mode image.
Perform the following steps to acquire PW-mode images.
1. Optimize a B-mode image.
2. Select the
PW
key on the control panel to enter the inactivated B+PW mode 1.
Содержание EVUS 8
Страница 1: ...C d Rev 02 77000001436 EVUS 8 OWNER S MANUAL English...
Страница 10: ...This page is intentionally left blank...
Страница 18: ...This page is intentionally left blank...
Страница 62: ...This page is intentionally left blank...
Страница 88: ...This page is intentionally left blank...
Страница 92: ...This page is intentionally left blank...
Страница 112: ...This page is intentionally left blank...
Страница 122: ...This page is intentionally left blank...
Страница 149: ...139 Appendix E Acoustic Output Data Please refer to Section 4 9 2 Acoustic Output...
Страница 150: ...NUM REG ANVISA 10069210070 www saevo com br...