4 Customizing Your System
Basic User Manual 39
Item
Description
Relate to CommitStore
Select the desired DICOM commitment storage server to be associated with the
DICOM storage server.
Send After Every
Image Stored
Tick this item and the system sends an image to the DICOM storage server every
time the image is saved.
Send At End of Exam
Tick this item and the system sends all saved images to the DICOM storage server
after the exam is complete.
Click Echo to verify the connection between the ultrasound system and the DICOM storage server after you input
the settings for the
Remote HostName
,
IP address
,
DICOM AE Title
,
Port Number
. A summary displays if the
connection is verified.
4�7�2
Commitment Settings
You can connect the ultrasound system to the DICOM commitment storage server and adjust the relevant settings
on the Commitment Storage tab.
Delete
Add
DICOM Storage Commitment
System Setting
Store
C-Store
Worklist
MPPS
Service List
StoreCommitment
Remote HostName
Synchronously Receive N-EVENT-REPORT Message
StoreCommitment
Port Number
105
Connect Timeout(sec)
10
Repeat Count
1
Dimse Timeout (sec)
10
Acse Timeout (sec) 30
Ping
Echo
IP Address
192.168.254.104
DICOM AE Title
AE
Peripheral
General
Comment
Measure
DICOM
Load Default
About
Exit
Save&Exit
Bodymark
Report
Figure 4-17 DICOM Settings Menu - Commitment Storage Tab
Item
Description
Service List
●
Select
Append
from the drop-down list beside
Service List
, input the relevant
parameters information and click
Add
to add a new server.
●
Select any server from the drop-down list and click
Delete
to delete the server.
Remote HostName
Enter the name of the DICOM commitment storage server.
IP Address
Set the IP address of the DICOM commitment storage server.
Click
Ping
to test the connection.
Содержание EVUS 8
Страница 1: ...C d Rev 02 77000001436 EVUS 8 OWNER S MANUAL English...
Страница 10: ...This page is intentionally left blank...
Страница 18: ...This page is intentionally left blank...
Страница 62: ...This page is intentionally left blank...
Страница 88: ...This page is intentionally left blank...
Страница 92: ...This page is intentionally left blank...
Страница 112: ...This page is intentionally left blank...
Страница 122: ...This page is intentionally left blank...
Страница 149: ...139 Appendix E Acoustic Output Data Please refer to Section 4 9 2 Acoustic Output...
Страница 150: ...NUM REG ANVISA 10069210070 www saevo com br...