6 Acquiring Images
66
Basic User Manual
0
5
0
10
10
5
Y axis:
Frequency
X axis: Time
S
Figure 6-9 M-Mode Imaging Screen
−
X-axis is the time scale.
−
Y-axis is the depth scale.
NOTE:
Once the M mode is activated, you can move the trackball to stop the M trace and adjust the M line. The
system continues tracing if the trackball is idle for more than 0.5s.
5. Optimize M-mode image. For details, refer to Section 6.4.3 Optimizing M-Mode Images.
6. Select the
M
key to exit.
6�4�2
Anatomical M-Mode
Anatomical M-mode is used for fetal cardiac applications. Anatomical M-mode can be used with phased array
probes when performing cardiac exams or convex probes when performing abdominal exams. In the anatomical
M-mode, the M line can be positioned perpendicular to the anatomical structure and be adjusted 360° even when
viewing motion patterns for irregular objects. It is used to study the ventricular function of the heart.
Perform the following steps to acquire anatomical M-mode images.
1. Enter the inactivated B+M mode, and select the
Menu
key. A menu pops up on the screen.
2. Select
Steer M
by using the trackball or the ↑ or ↓ key.
3. Rotate the
Menu
key or select the ← or → key to set the number of M lines.
Multiple M lines display after selecting the desired number as the following figure shows.
0
5
10
M Line
2D Image
S
Figure 6-10
Inactivated Anatomical M-Mode Imaging Screen
Содержание EVUS 8
Страница 1: ...C d Rev 02 77000001436 EVUS 8 OWNER S MANUAL English...
Страница 10: ...This page is intentionally left blank...
Страница 18: ...This page is intentionally left blank...
Страница 62: ...This page is intentionally left blank...
Страница 88: ...This page is intentionally left blank...
Страница 92: ...This page is intentionally left blank...
Страница 112: ...This page is intentionally left blank...
Страница 122: ...This page is intentionally left blank...
Страница 149: ...139 Appendix E Acoustic Output Data Please refer to Section 4 9 2 Acoustic Output...
Страница 150: ...NUM REG ANVISA 10069210070 www saevo com br...