NB-IoT NPRACH Measurement
R&S
®
CMW-KM300/-KS300
365
User Manual 1178.5689.02 ─ 06
CONFigure:NIOT:MEAS<i>:PRACh:REPetition
<Repetition>
Specifies the repetition mode of the measurement. The repetition mode specifies
whether the measurement is stopped after a single shot or repeated continuously. Use
CONFigure:..:MEAS<i>:...:SCOunt
to determine the number of measurement
intervals per single shot.
Parameters:
<Repetition>
SINGleshot | CONTinuous
SINGleshot:
Single-shot measurement
CONTinuous:
Continuous measurement
*RST:
SING
Example:
See
Single-Shot and Continuous Measurements
Firmware/Software:
V3.7.20
Manual operation:
See
CONFigure:NIOT:MEAS<i>:PRACh:SCONdition
<StopCondition>
Qualifies whether the measurement is stopped after a failed limit check or continued.
SLFail
means that the measurement is stopped and reaches the
RDY
state when one
of the results exceeds the limits.
Parameters:
<StopCondition>
NONE | SLFail
NONE:
Continue measurement irrespective of the limit check
SLFail:
Stop measurement on limit failure
*RST:
NONE
Example:
See
Specifying Measurement-Specific Settings
Firmware/Software:
V3.7.20
Manual operation:
See
CONFigure:NIOT:MEAS<i>:PRACh:MOEXception
<MeasOnException>
Specifies whether measurement results that the R&S
CMW identifies as faulty or inac-
curate are rejected.
Parameters:
<MeasOnException> OFF | ON
OFF:
Faulty results are rejected
ON:
Results are never rejected
*RST:
OFF
Example:
See
Specifying Measurement-Specific Settings
Firmware/Software:
V3.7.20
Manual operation:
See
Command Reference