NB-IoT Multi-Evaluation Measurement
R&S
®
CMW-KM300/-KS300
309
User Manual 1178.5689.02 ─ 06
Firmware/Software:
V3.5.10
Table 3-4: Trace structure depending on subcarrier spacing
SC spacing
Covered time
Value spacing
3.75 kHz
-2200 μs to 2196.875 μs
3.125 μs
15 kHz
-1100 μs to 1098.4375 μs
1.5625 μs
CALCulate:NIOT:MEAS<i>:MEValuation:PDYNamics:CURRent?
CALCulate:NIOT:MEAS<i>:MEValuation:PDYNamics:AVERage?
CALCulate:NIOT:MEAS<i>:MEValuation:PDYNamics:MINimum?
CALCulate:NIOT:MEAS<i>:MEValuation:PDYNamics:MAXimum?
FETCh:NIOT:MEAS<i>:MEValuation:PDYNamics:CURRent?
FETCh:NIOT:MEAS<i>:MEValuation:PDYNamics:AVERage?
FETCh:NIOT:MEAS<i>:MEValuation:PDYNamics:MINimum?
FETCh:NIOT:MEAS<i>:MEValuation:PDYNamics:MAXimum?
FETCh:NIOT:MEAS<i>:MEValuation:PDYNamics:SDEViation?
READ:NIOT:MEAS<i>:MEValuation:PDYNamics:CURRent?
READ:NIOT:MEAS<i>:MEValuation:PDYNamics:AVERage?
READ:NIOT:MEAS<i>:MEValuation:PDYNamics:MINimum?
READ:NIOT:MEAS<i>:MEValuation:PDYNamics:MAXimum?
READ:NIOT:MEAS<i>:MEValuation:PDYNamics:SDEViation?
Return the current, average, minimum, maximum and standard deviation single value
results of the power dynamics measurement.
The values described below are returned by
FETCh
and
READ
commands.
CALCulate
commands return limit check results instead, one value for each result listed below.
Return values:
<Reliability>
<OutOfTolerance>
Out of tolerance result, i.e. percentage of measurement inter-
vals of the statistic count for power dynamics measurements
exceeding the specified power dynamics limits.
Default unit: %
<OffPowerBefore>
Default unit: dBm
<OnPowerRMS>
Default unit: dBm
<OnPowerPeak>
Default unit: dBm
<OffPowerAfter>
Default unit: dBm
Example:
See
Usage:
Query only
Firmware/Software:
V3.5.10
Command Reference
深圳德标仪器