NB-IoT Multi-Evaluation Measurement
R&S
®
CMW-KM300/-KS300
234
User Manual 1178.5689.02 ─ 06
Figure 3-18: Measurement control settings - part 2
Modulation Scheme < Modulation
...............................................................................235
EVM Excl. Periods Leading/Lagging < Modulation
..................................................... 236
.............................................................................. 236
.................................................................................236
Offline Segment Nr. < List Mode
................................................................................. 236
Physical Cell ID
Adjust this parameter to the physical cell ID used by the measured signal. This setting
is required to ensure that the R&S
CMW can synchronize to the signal and perform
channel estimation.
In the standalone (SA) scenario, this parameter is controlled by the measurement. In
the combined signal path (CSP) scenario, it is controlled by the signaling application.
Remote command:
CONFigure:NIOT:MEAS<i>:MEValuation:PLCid
(SA)
CONFigure:NIOT:SIGN<i>:CELL:PCID
(CSP)
Delta Seq. Shift
Delta sequence shift value (Δ
ss
) used to calculate the sequence shift pattern for the
NPUSCH. Adjust this value to the measured signal to ensure that the R&S
CMW can
synchronize to the signal and perform channel estimation.
Remote command:
CONFigure:NIOT:MEAS<i>:MEValuation:DSS
GUI Reference