NB-IoT Multi-Evaluation Measurement
R&S
®
CMW-KM300/-KS300
214
User Manual 1178.5689.02 ─ 06
Figure 3-6: Error Vector Magnitude view
●
Upper diagram
The bar graph shows the EVM for each SC-FDMA symbol in the measured slot.
The symbols are numbered from 0 to 6, with the reference symbol labeled 3.
Each value is an RMS value or a peak value over all modulation symbols in the
SC-FDMA symbol. The title above the bar graph indicates whether RMS or peak
values are currently displayed. To switch between RMS and peak values, use the
softkey - hotkey combination "Display" - "Select Trace".
●
Lower diagram
The lower diagram shows the EVM for all modulation symbols in the measured
slot. The results in the upper diagram are calculated from these data.
For a description of the statistical overview, see
Chapter 3.3.2.10, "View TX Measure-
For the parameters at the bottom, see
Chapter 3.3.2.13, "Common View Elements"
For query of the diagram contents via remote control, see
3.3.2.3
Views Magnitude Error, Phase Error
Both views show a diagram and a statistical overview of results per slot.
GUI Reference
深圳德标仪器