CONFigure:NIOT:MEAS<i>:MEValuation:LIMit:SEMask:LIMit<no>
<Enable>,
<FrequencyStart>, <FrequencyStop>, <PowerLevelStart>, <PowerLevelStop>
Defines the emission mask area <no>.
Suffix:
<no>
.
1..12
Number of the emission mask area
Parameters:
<Enable>
OFF | ON
OFF
: disables the check of these requirements
ON
: enables the check of these requirements
*RST:
ON for area 1 to 5, OFF for area 6 to 12
<FrequencyStart>
Start frequency of the area, relative to the edges of the channel
bandwidth
Range:
0 MHz to 1.9 MHz
*RST:
depends on the area number
Default unit: Hz
<FrequencyStop>
Stop frequency of the area, relative to the edges of the channel
bandwidth
Range:
0 MHz to 1.9 MHz
*RST:
depends on the area number
Default unit: Hz
<PowerLevelStart>
Upper limit at the <FrequencyStart>
Range:
-256 dBm to 256 dBm
*RST:
depends on the area number
Default unit: dBm
<PowerLevelStop>
Upper limit at the <FrequencyEnd>
Range:
-256 dBm to 256 dBm
*RST:
depends on the area number
Default unit: dBm