NB-IoT Multi-Evaluation Measurement
R&S
®
CMW-KM300/-KS300
222
User Manual 1178.5689.02 ─ 06
Figure 3-14: Power Dynamics view
There are two diagrams. One shows the start of the first allocated resource unit (RU).
The other one shows the end of the last allocated RU.
The covered time intervals shown on the X-axis are as follows:
●
The origin 0 µs marks the start of the first allocated RU in the left diagram and the
end of the last allocated RU in the right diagram.
●
For 3.75 kHz SC spacing, the axis covers -2200 µs to 2200 µs.
●
For 15 kHz SC spacing, the axis covers -1100 µs to 1100 µs.
The individual ON and OFF power areas are labeled in the diagrams. Exclusion peri-
ods are marked by gray vertical bars.
The table below the diagrams shows a statistical evaluation of the ON power and OFF
power values. For a definition of the values, see
Chapter 3.3.6.8, "Power Dynamics
For the parameters at the bottom of the view, see
Chapter 3.3.2.13, "Common View
For query of the diagram and table contents via remote control, see
GUI Reference