NB-IoT Signaling
R&S
®
CMW-KM300/-KS300
189
User Manual 1178.5689.02 ─ 06
INITiate:NIOT:SIGN<i>:BLER
STOP:NIOT:SIGN<i>:BLER
ABORt:NIOT:SIGN<i>:BLER
Starts, stops, or aborts the measurement:
●
INITiate...
starts or restarts the measurement. The measurement enters the
"RUN" state.
●
STOP...
halts the measurement immediately. The measurement enters the "RDY"
state. Measurement results are kept. The resources remain allocated to the mea-
surement.
●
ABORt...
halts the measurement immediately. The measurement enters the
"OFF" state. All measurement values are set to NAV. Allocated resources are
released.
Use
FETCh...STATe?
to query the current measurement state.
Usage:
Event
Firmware/Software:
V3.7.20
Manual operation:
See
FETCh:NIOT:SIGN<i>:BLER:STATe?
Queries the main measurement state. Use
FETCh:...:STATe:ALL?
to query the
measurement state including the substates. Use
INITiate..., STOP...,
ABORt...
to change the measurement state.
Return values:
<State>
OFF | RUN | RDY
OFF:
measurement off, no resources allocated, no results
RUN:
measurement running, synchronization pending or adjus-
ted, resources active or queued
RDY:
measurement terminated, valid results can be available
*RST:
OFF
Usage:
Query only
Firmware/Software:
V3.7.20
Manual operation:
See
FETCh:NIOT:SIGN<i>:BLER:STATe:ALL?
Queries the main measurement state and the measurement substates. Both measure-
ment substates are relevant for running measurements only. Use
FETCh:...:STATe?
to query the main measurement state only. Use
INITiate..., STOP...,
ABORt...
to change the measurement state.
Command Reference