NB-IoT Multi-Evaluation Measurement
R&S
®
CMW-KM300/-KS300
245
User Manual 1178.5689.02 ─ 06
The following figure provides an overview of the relevant time intervals. The power
dynamics measurement determines the ON power, the OFF power (before) and the
OFF power (after), using the indicated time intervals.
Figure 3-27: OFF/ON transition, start of allocation
Figure 3-28: ON/OFF transition, end of allocation
The OFF power must not exceed -48.5 dBm. The ON power must be between -18.5
dBm and -3.5 dBm.
Characteristics
Refer to 3GPP TS 36.521 V14.4.0, section...
Specified limit
ON power
6.3.4F.1 General ON/OFF time mask for category NB1
≥ -18.5 dBm, ≤ -3.5
dBm
OFF power
6.3.4F.1 General ON/OFF time mask for category NB1
≤ -48.5 dBm
Power dynamics limit settings
The limits can be set in the configuration dialog.
Figure 3-29: Power dynamics limit settings
Remote command:
CONFigure:NIOT:MEAS<i>:MEValuation:LIMit:PDYNamics
GUI Reference