Fitting the Options
SMIQ
1125.5555.03
E-7
1.10
1.3.8
Option SMIQB15 - Second Fading Simulator (FSIM2)
The second Fading Simulator is fitted at the slot with label ’FSIM2’.
Before fitting the option SMIQB15 (FSIM2) please check the correct settings of the jumpers on both
fading modules FSIM1 (see Section 1.3.7) and FSIM2. The jumper settings depend on the series of the
FSIM module, they are also indicated on the screening cover.
After an instrument warm-up period of 2 hours, the IQ Modulator should be calibrated.
Checking the jumper
Jumper Setting for the Fading Simulator FSIM2:
settings
(cf. module number and labeling ’Jumper Setting’ on the screening
cover of the module)
FSIM2
FADING SIMULATOR
1085.XXXX
FADING SIMULATOR
1114.XXXX
Jumper connects
X1.2
with
X1.3
X1.1
with
X1.2
X8.2
with
X8.3
X8.2
with
X8.3
X9.2
with
X9.3
X9.1
with
X9.2
X12.1
with
X12.2
X12.1
with
X12.2
X13.2
with
X13.3
X13.2
with
X13.3
X15.2
with
X15.3
X15.2
with
X15.3
X16.2
with
X16.3
none
Fading modules of series 1114.XXXX also require the setting of a DIP switch. The setting is indicated
on the screening cover.
FADING SIMULATOR
1114.XXXX
Switch
1
2
FSIM1
OFF
ON
FSIM2
ON
OFF