Bit Error Rate Test
SMIQ
1125.5555.03
2.372
E-9
PATTERN IGNORE
In the event of a so-called frame error (eg an error detected in the check sum),
some mobile radio standards communicate a complete "0" or "1" frame to the
BERT instead of the detected frame data. This is to signal to the BERT that
this frame is not to be used for the BER measurement.
OFF
PATTERN IGNORE is not active.
ALL 1
Bit sequences consisting of 31 or more subsequent "1" data
are not used (ie ignored) for the BER measurement.
ALL 0
Bit sequences consisting of 31 or more subsequent "0" data
are not used (ie ignored) for the BER measurement.
IEC/IEEE-bus command
:BERT:SET:IGN OFF
Notes
:
-
While ignoring the bits, the PRBS generator for the BER measurement
keeps running. Following the "0" or "1" sequence, the BER measurement
is continued as if the ignored sequence had contained PRBS data.
-
If PATTERN IGNORE is switched on, synchronization time is 32 bits
longer.
-
If PATTERN IGNORE and RESTART are active at the same time, the
presence of the restart signal causes the measurement to stop
immediately
. None of the 32 bits within the PATTERN IGNORE detector
is evaluated.
-
If DATA ENABLE and PATTERN IGNORE are active at the same time,
DATA ENABLE is given priority, ie bits that are not enabled are not
examined for "0" or "1" sequences.
50 Bits
PRBS at Generator
PRBS at DUT Output
BER "active"
Drop Out of DUT
These bits are ingored
Example
50 bits were set to "0" by the DUT. These 50 bits plus the preceding
"0" are ignored in the measurement.
UNIT
Setting the unit for displaying the bit error rate.
OFF
No unit, decade values
%
Relative frequency in percent
ppm
Relative frequency in parts per million
IEC/IEEE-bus command
:BERT:UNIT OFF