External Modulation Source AMIQ
SMIQ
1125.5555.03
2.360
E-9
(SETUP...)
OFFSET I VAR
Fine adjustment of level offset for the I channel in
case of variable level. The offset detuning depends
on the of the mechanical attenuator setting of AMIQ.
See AMIQ manual, section SOURce - Hardware
Settings.
LEVEL Q GAIN
Fine adjustment of gain factor for the Q channel in
case of fixed level.
OFFSET Q FIX
Fine adjustment of level offset for the Q channel in
case of fixed level.
OFFSET Q VAR
Fine adjustment of level offset for the Q channel in
case of variable level.
SKEW (Q-I)
Defining the delay between the I and Q channel.
Positive values delay I compared with Q. The delay
can be modified by entering the relative factors
±
1.0
in a range of approx.
±
1 ns at a resolution of 10 ps.
SELF TEST
Triggers the internal AMIQ selftest which comprises
the following:
−
memory test of the output memory.
−
control of components of analog hardware and
check of signal paths by means of the built-in
diagnostic A/D converter.
SAVE/RECALL
SETTINGS...
Opens a window to save/load/delete a maximum number of 100 AMIQ setups.
SETUP...
SAVE/RECALL
SELECT WAVE
MODE
EXECUTE SIN
EXT TRIG SL
FILTER
LEVEL...
MARKER...
REFERENCE O
FREQ
100. 000 000 0
MHz
L EV E L
- 30.0
d Bm
F RE Q UE N CY
L EV E L
A NA L OG MO D
V EC T OR MO D
D IG I TA L M O D
D IG I TA L S T D
A MI Q C T RL
L F O UT P UT
S WE E P
L IS T
M EM SE Q
RECALL SETTING...
SAVE SETTING...
DELETE SETTING...
Fig. 2-229
Menu AMIQ CTRL -SAVE/RECALL SETTINGS...
(SAVE/RECALL
SETTINGS...)
RECALL SETTING...
Selects and loads a stored AMIQ setup. The
available setups are offered in a select window.