Digital Standard PDC
SMIQ
1125.5555.03
2.284
E-9
(MODULATION...)
SET TO
STANDARD
Sets the subsequent modulation parameters to the
values predefined by the standard.
MODULATION TYPE
Displays the modulation type.
SYMBOL RATE
Input value for the symbol clock. 21 ksymbol/s are
preset.
IEC/IEEE-bus
:SOUR:PDC:SRAT 21.1KHZ
FILTER
Selection of baseband filter. A selection between
Nyquist filters COS, SQRCOS or a user-defined filter
USER (cf. Section Digital Modulation) is possible.
IEC/IEEE-bus
:SOUR:PDC:FILT:TYPE COS
ROLL OFF FACTOR
Input value for the roll-off factor.
IEC/IEEE-bus
:SOUR:PDC:FILT:PAR 0.51
FILTER MODE
Selection of filter mode.
LOW_ACP Filter for minimum Adjacent Channel
Power.
IEC
:SOUR:PDC:FILT:MODE LACP
LOW_EVM Filter for minimum vector error.
IEC
:SOUR:PDC:FILT:MODE LEVM
LOW DISTORTION
Switch on/off of low-distortion mode..
After switch-on, the level of the IQ baseband signals
is reduced by 3 dB. In some cases, this might reduce
undesired intermodulation products. OFF is normally
the more favourable setting.
IEC/IEEE-bus command
:SOUR:PDC:LDIS OFF
TRIGGER MODE...
Opens a window for selecting the trigger mode.
AUTO
The PDC signals are continuously transmitted in the
activated slots.
IEC/IEEE-bus
:SOUR:PDC:SEQ AUTO
RETRIG
The PDC signals are continuously transmitted in the
activated slots. A trigger event causes a restart.
IEC/IEEE-bus
:SOUR:PDC:SEQ RETR
ARMED_AUTO
The PDC signal generation does not start until a
trigger event has occurred. The unit then
automatically switches over to the AUTO mode and
can no longer be triggered.
IEC/IEEE-bus
:SOUR:PDC:SEQ AAUT
ARMED_RETRIG
The PDC signal generation does not start until a
trigger event has occurred. The unit then
automatically switches over to the RETRIG mode.
Each new trigger event causes a restart.
IEC/IEEE-bus
:SOUR:PDC:SEQ ARET