SMIQ
RF Level
1125.5555.03
E-8
2.49
EXCLUDE FROM RCL
OFF
The saved RF level is also loaded when instrument settings are
loaded with the [RCL] key or with a memory sequence.
IEC/IEEE-bus command
SOUR:POW:RCL INCL
ON
The RF level is not loaded when instrument settings are loaded,
the current level is maintained.
IEC/IEEE-bus command
SOUR:POW:RCL EXCL
2.5.1 Level
Offset
The SMIQ offers the possibility of entering the offset (OFFSET) of a possibly series-connected
attenuator/amplification element in the LEVEL-LEVEL menu. The indication/input value in the LEVEL
header field considers this input (see below) and represents the level value of the signal at the output of
the series-connected instrument (cf. Fig. 2-19).
The entry values AMPLITUDE and OFFSET in the menu LEVEL or LEVEL in the header line are related
to the RF output level as follows:
LEVEL - OFFSET = output level (= AMPLITUDE)
An offset entry causes no modification of the RF output frequency but only a modification of the display
value in the LEVEL field in the header line, ie LEVEL in the header line indicates the offset-associated
level and AMPLITUDE in the menu LEVEL the RF output level. The status line indicates LEV-OFFST.
The offset is to be entered in dB.
The offset setting also remains effective in the ATTENUATOR MODE FIXED operating mode and with
level sweep.
ï ð
SMIQ
Input value
OFFSET
(LEVEL menu)
RF output signal
(AMPLITUDE in the LEVEL menu)
Attenuation/
Amplification
LEVEL in the header field
Fig. 2-19
Example of a circuit with level offset