Arbitrary Waveform Generator ARB
SMIQ
1125.5555.03
E-9
2.348
(TRIGGER...)
TRIGGER OUT1(2) MODE
The user may select one of the displayed modes for
generation of the trigger output signals.
Depending on the system, WinIQSIM can calculate clock
signals for BIT, SYMBOL, SLOT, FRAME and the restart
signal (sequence) and transmit them together with the
waveform. These modes set the counters for ON TIME
and OFF TIME (see below). The names are defined as
modes by WinIQSIM, stored in the waveform where they
can be selected. The maximum number of defined
counters is 4. It may occur that only the SEQUENZ and
USER modes can be provided since the entirety of modes
is not always available for all systems.
BIT_CLK
edge at beginning of each bit
SYMB_CLK
edge at beginning of each modulation
symbol
SLOT_CLK
edge at beginning of each slot
FRAM_CLK
edge at beginning of each frame
SEQUENZ
edge at beginning of each sequence
(may, e.g., be used as restart signal for
BERT)
PULSE
pulse defined by WinIQSIM user
USER
The user defines a pulse by means of
the parameters for ON and OFF TIME.
IEC/IEEE bus command
:ARB:TRIG:OUTP:MODE USER
....ON TIME
Length of the active state of the output trigger signals
indicated as number of samples. Can be modified only, if
MODE = USER.
This time is repeated after OFF TIME has run off.
IEC/IEEE bus command
:ARB:TRIG:OUTP:ONT 3789
....OFF TIME
Length of the non-active state of the output trigger signals
indicated as number of samples. Can be modified only, if
MODE = USER.
IEC/IEEE bus command
:ARB:TRIG:OUTP2:OFFT 639