Digital Standard GPS
SMIQ
1125.5555.03
E-9
2.336
SET RF AND POWER
DEFAULT
Sets the RF default setting for GPS:
•
The frequency FREQ is set to the GPS carrier frequency L1 of 1.57542
GHz.
•
The output level LEVEL is set to –125.0 dBm.
IEC/IEEE-bus command
:SOUR:GPS:PRES:RF
TRIGGER MODE
Configuration of autorun control of generated GPS signal.
AUTO
The GPS signal calculated in realtime is immediately output and
cyclically repeated. Trigger events are ignored.
IEC/IEEE-bus command
:SOUR:GPS:SEQ AUTO
RETRIG
The GPS signal calculated in realtime is immediately output and
cyclically repeated. A trigger event causes a restart from
the first navigation data bit.
IEC/IEEE-bus command
:SOUR:GPS:SEQ RETR
ARMED_AUTO
Only a trigger event causes a start of the GPS signal from
the first navigation data bit. Further trigger events are
ignored.
IEC/IEEE-bus command
:SOUR:GPS:SEQ AAUT
ARMED_RETRIG Only a trigger event causes a start of the GPS signal from
the first navigation data bit. Any other trigger event causes
a restart.
IEC/IEEE-bus command
:SOUR:GPS:SEQ ARET
EXECUTE TRIGGER
Executes a trigger event to start the GPS signal.
IEC/IEEE-bus command
:TRIG:DM:IMM
TRIGGER...
Opens a window for selecting the trigger source, configuring trigger output
signals, and setting the delay of an external trigger signal.
TRIGGER SOURCE
Selects the trigger source.
EXT
The GPS signal is started from the first
navigation bit with the active edge of an
external trigger signal. The polarity, trigger
threshold, and input impedance of the
TRIGIN input can be changed in the DIGITAL
MOD - EXT INPUTS menu.
INT
A trigger event is manually started by
EXECUTE TRIGGER.
IEC/IEEE-bus command
:SOUR:GPS:TRIG:SOUR
EXT
EXT TRIGGER
DELAY
Sets the number of chips by which an external trigger
signal is delayed before it starts the GPS signal. This
is used to set up synchronization with the DUT or
other units.
IEC/IEEE-bus command
:SOUR:GPS:TRIG:DEL 3