SMIQ
Digital Standard DECT
1125.5555.03
2.319
E-9
2.19.1
Sync and Trigger Signals
The data generator generates a data sequence with modulation data, control signals for envelope
control, and synchronization signals.
When TRIGGER MODE AUTO is selected, the DECT signal generation starts automatically.
This start can also be activated by an external trigger signal (TRIGGER MODE ARMED_AUTO) which
allows a synchronous sequence for BER measurements to be carried out on receivers.
Trigger signals for synchronized sequences can be used for measuring the bit error rate of receivers. A
trigger signal can be fed via the TRIGIN input at connector PAR DATA. The active slope of a trigger
signal applied there executes a trigger event.
DECT signal generation at a frame limit is started after a trigger event. Data from data lists are inserted
into the selected slots starting from the first bit. PRBS generators start with the set initialization status.
Signal generation either starts immediately after the active slope of the trigger signal or after a settable
number of symbols (EXT TRIGGER DELAY). Retriggering (RETRIG) can be inhibited for a settable
number of symbols (EXT RETRIGGER INHIBIT).
A trigger event can be executed manually or via the IEC/IEEE bus using EXECUTE TRIGGER.
When a trigger event is executed, a trigger signal is output at the TRIGOUT 3 output of SMIQ.
SMIQ also generates the following sync signals:
•
a frame clock at TRIGOUT 1 output,
•
a frame or multiframe clock at TRIGOUT 2 output with settable position in the frame,
•
the symbol clock and the bit clock.
A clock synthesizer on the modulation coder generates the symbol clock and the bit clock in SMIQ. All
clock signals are synchronized to the 10-MHz reference of SMIQ. The bit clock is available at connector
BIT CLOCK. If desired, the clock synthesizer in SMIQ can be synchronized to an external bit clock.
The clock source is selected in the CLOCK-CLOCK SOURCE EXT menu.
To allow for a trouble-free synchronization of the clock synthesizer first apply the external clock and set
the correct symbol rate at SMIQ. Then switch CLOCK SOURCE from INT to EXT.
Note:
The set symbol rate should not differ by more than 1% from the symbol rate of the external
signal.