SMIQ
Functional Test
1125.5555.03 E-7
1.3
1.1.6 Preset Setting
A defined setting status is achieved by pressing the [PRESET] key.
Preset Status:
RF frequency 100 MHz
RF level -30 dBm
Reference frequency internal, adjustment off
Offsets 0
Modulations switched off
Transient-free level setting switched off, level attenuator mode: AUTO
Internal level control level ALC: AUTO
User correction
level UCOR: OFF
LF output switched off
Sweep switched off
List mode switched off
Memory sequence switched off
Suppression of indications system security: unaltered
Protection of calibration data protection lock: unaltered
Settings stored unaltered
Data, lists etc. stored unaltered
IEC-bus address unaltered
Beeper unaltered
All parameters and circuit states, even those of operating modes which are not activated, are preset by
means of Preset.
The presettings going beyond the above list can be seen from the menu representations as of Section
2.4 which each indicate the Preset setting status.
1.2 Functional Test
On switching on the instrument and permanently during operation, the SMIQ carries out a self test. The
ROM contents as well as the battery of the non-volatile RAM are checked on switching on the
instrument and the RAM contents with every calling the memory. The most important instrument
functions are automatically monitored during operation.
If an error is detected, the message "ERROR" is displayed in the status line. For further identification of
the error, press the [ERROR] key. Thereupon a description of the error/s is displayed (cf. Chapter 2,
section "Error Messages"). Return to the menu exited by pressing the [RETURN] key.
If required, the self tests can be induced purposefully. See Chapter 4, section "Functional Test".
Further, internal test points can be polled by the user and the results be read out and displayed. See
Chapter 2, section "Voltage Indication of Test Points".