LIST Mode
SMIQ
1125.5555.03
2.404
E-9
MODE
Selection of the operating mode.
IEC/IEEE-bus commands :
:SOUR:FREQ:MODE LIST;
:SOUR:LIST:MODE AUTO;
:TRIG:LIST:SOUR SING
EXECUTE SINGLE LIST
Starts a single run of a list. This menu option is only visible if MODE
SINGLE is selected.
IEC/IEEE-bus command :
:TRIG:LIST
RESET LIST
Sets the starting point.
IEC/IEEE-bus command
:ABOR:LIST
DWELL
Input value of the dwell time per step.
IEC/IEEE-bus command
:
SOUR:
LIST:DWEL 10ms
CURRENT INDEX
Indication of the current list index. Setting value of the current list index
in the STEP operating mode.
LEARN
Starts the LEARN function. All value pairs of the active list are
subsequently set by the instrument with the current additional
parameters, and the hardware setting data are stored.
Caution:
This function must be called after every creating and
altering the list (or the remaining setting data). The list
must be learned as well after temperature changes or
after calling internal calibration routines, as all parameters
usually controlled are replaced by stored values in the
LIST mode.
IEC/IEEE-bus command
:SOUR:LIST:LEAR
SELECT LIST...
Selection of a list or creation of a new list (cf. Section 2.2.11, List
Editor).
IEC/IEEE-bus command
:SOUR:LIST:SEL 'LIST2'
DELETE LIST...
Deletion of a list (cf. Section 2.2.11, List Editor).
IEC/IEEE-bus command
:SOUR:LIST:DEL
'LIST1'
FUNCTION
Selection of the editor functions to process a list (cf. Section 2.2.11,
List Editor).
IEC/IEEE
:SOUR:LIST:FREQ 100MHz, 1.2GHz; POW 0dBm, 6dBm