Digital Standard NADC
SMIQ
1125.5555.03
E-9
2.270
(CLOCK...)
CLOCK SOURCE
Selection of clock source.
INT
SMIQ uses internally generated clock signals.
EXT
An external symbol clock or bit clock is fed in
at connectors SYMBOL CLOCK or BIT
CLOCK. The clock synthesizer on the
modulation coder is synchronized to this
clock.
The symbol rate has to be set with an
accuracy of
±
1 %.
The polarity, the trigger threshold and the
input resistance of the clock inputs can be
modified in menu DIGITAL MOD - EXT
INPUTS.
IEC/IEEE-bus command
:SOUR:NADC:CLOC:SOUR INT
MODE
Selection of clock for external clock signal.
SYMBOL
The external clock has to be a symbol clock.
BIT
The external clock has to be a bit clock.
IEC/IEEE-bus command
:SOUR:NADC:CLOC:MODE SYMB
DELAY
Setting the delay of generated modulation signal to an
external clock.
This can be used, for example, for synchronization with a
second unit to achieve time synchronization between the
modulation signals of the two units.
IEC/IEEE-bus command
:SOUR:NADC:CLOC:DEL 0.5
POWER RAMP
CONTROL...
Opens a window for setting the envelope control, especially for the rising and
falling ramp at the beginning and end of a slot.
FREQ
100. 000 000 0
MHz
FREQUENCY
LEVEL
ANALOG MOD
VECTOR MOD
DIGITAL MOD
DIGITAL STD
LF OUTPUT
SWEEP
LIST
MEM SEQ
UTILITIES
NADC
PHS
IS95
NADC
PDC
GSM
STATE
MODULATI ON...
TRIGGER MODE
EXECUTE TRIG
TRIGGER. ...
CLOCK...
POWER RA MP C>
SLOT ATT ENUA
1
TCH
SET DEFAULT
RAMP TIME
2.0 Symb
RAMP FUNCTION
LIN
COS
RAMP DELAY
0.0 Symb
RISE OFFSET
0 Symb
FALL OFFSET
0 Symb
LEVEL
- 30.0
dBm
PEP
- 27.6
dBm
Fig. 2-180
Menu DIGITAL STD - NADC - POWER RAMP CONTROL... , SMIQ equipped with
Modulation Coder SMIQB20 and Data Generator SMIQB11