TROUBLESHOOTING
8 - 11
Problem
Possible Reasons & Solutions
Reading
abnormally
high
Reasons:
Wrong calibration data.
Dirty sensor module.
Dirty water trap filter.
Excessive moisture and water
condensation.
Solutions:
Re-calibrate with the gas.
Clean sensor module.
Replace water trap filter.
Blow dry the sensor module.
“Lamp”
message
during
operation
Reasons:
Wrong threshold.
Dirty PID sensor.
Weak or defective PID lamp.
Solutions:
Adjust the lamp threshold.
Clean PID sensor.
Replace PID lamp.
Reading
abnormally
low
Reasons:
Incorrect calibration.
Low sensitivity to the specific
gas.
Solutions:
Calibrate the monitor.
Replace sensor.