PROGRAMMING OF MULTIRAE
4 - 18
4.4.5 Change LEL/VOC Span Gas
This function allows the user to select a specific LEL or
VOC gas to be used as the span gas during LEL or VOC gas
calibration.
1. “Change LEL/VOC Span Gas?” is the next sub-menu item
in Table 4.3.
2. Press
[Y/+]
key. If LEL sensor is installed and enabled,
the display shows :
TOX1
VOC
TOX2
LEL Span = ?
methane
LEL
OXY
Otherwise, the message “No LEL installed” will appear.
3. If the user does not want to change the LEL span gas,
press
[Y/+]
key to accept the current selection and exit
this sub-menu.
4. If the user wants to select a different LEL span gas, press
[N/-]
key. Then use
[Y/+]
or
[N/-]
key to scroll through
a list of gas names until a desired gas name appears in the
LCD display, then press
[MODE]
key to select the new
gas name.
5. The display shows “Save ?” To confirm the new gas
selection, press the
[Y/+]
key to accept the change. Press
the
[N/-]
key or the
[MODE]
key to discard the change
and move to the next display.
6. If the VOC sensor is installed and enabled, the display
shows:
TOX1
VOC
TOX2