![Rae MultiRAE PLUS PGM-50 Скачать руководство пользователя страница 108](http://html1.mh-extra.com/html/rae/multirae-plus-pgm-50/multirae-plus-pgm-50_operation-and-maintenance-manual_3369931108.webp)
COMPUTER INTERFACE FOR MULTIRAE PLUS
5 - 12
Note
: The VOC gas configuration settings applies only to
monitors which are equipped with PID detector option.
The
General
tab page, which is shown in Figure 5-7, allows
a user to change the general settings of the configuration
data. The following is the description of each field in the
General
tab page of the
Edit Configuration File
dialog:
ID:
This section sets the identification that is used throughout
the datalog process. Both editing fields,
Site
and
User,
can
hold up to eight alphanumeric letters.
Security Level:
There are three levels of security. They are used to provide
protection against unauthorized changes to monitor settings
which can be changed by using the programming menu of the
MultiRAE PLUS monitor when the instrument is in the
programming mode.
Level 0
allows a user to enter the programming menu
without password checking, but does not allow changes to
alarm limits, real time clock, correction factor, or user and
site information, etc. The user can still perform calibrations.
If
Level 1
is selected, a four digit password is required to
enter the programming menu.
If
Level 2
is selected, the programming menu may be
entered without restriction. Any changes made in the
programming menu can be saved.