COMPUTER INTERFACE FOR MULTIRAE PLUS
5 - 23
5.5.4 Configuring All Settings
A user can use this function to update the entire
configuration settings (except the instrument ID, serial
number and the datalog option) in the MultiRAE monitor.
To configure all the settings in the instrument, click the
Option->Config All
menu item, and a message box will
appear to remind the user to connect the instrument to the
PC through the serial port. After making sure that the
instrument is connected to the PC, click the
OK
button.
Another message box will appear to warn the user that this
operation will erase all the current configuration settings in
the monitor. If the user does not want this to happen, click
the
Cancel
button, otherwise click the
OK
button to start
the data transfer.
After the information is sent successfully, a message box
will appear to indicate that the sending configuration process
is completed. Click the
OK
button to close the message
box.
Warning
:
The
Config All
option is mainly used to restore
the MultiRAE monitor’s default operational settings when
the configuration of the unit is non-recoverable and the unit
is malfunctioning. Because the default parameters are not
normally fit a specific unit, a user will need to re-
calibration, re-setup lamp, pump, and LCD contrast
threshold, etc. Try not to use this function if possible.