PROGRAMMING OF MULTIRAE
4 - 14
4.4.3 Single Sensor Calibration
This procedure determines the second point of the sensor
calibration curve for a single sensor. A standard reference
gas (span gas) is needed to perform this procedure. Table
2.2 shows the standard calibration gas which is typically
used as the span gas in the factory (connect the tube to the
sampling bag filled with the standard gas as shown in Figure
4-1).
1. Continuing from Step 4 or Step 6 of the previous
section, the display should show “Single Sensor
Calibration?” Press
[Y/+]
key. The display shows all the
installed sensors in the monitor with the cursor blinking
at the first sensor location. Press
[Y/+]
key to select the
high-lighted sensor and start the calibration, or press
[MODE]
key momentarily to move to the next sensor
location.
TOX1
VOC
TOX2
CO
VOC
H
2
S
LEL pick?
OXY
LEL
OXY
2. Turn on the valve of the CO gas bottle to start the flow
of the span gas. Display shows “Apply CO Gas” and will
wait for the calibration gas to reach the sensor. When
the calibration gas has reached the sensor the display
will show “calibration in progress…60” with the
countdown timer shows the number of remaining
seconds while the monitor performs calibration. When