TROUBLESHOOTING
8 - 9
11) Check Back Light Threshold
This display can be used to adjust the LCD back light turn on
and off threshold. Verify that the threshold is 100 which
will fit almost all of the situation.
12) Serial Number and Pump Toggle
This display shows the unit’s serial number. If a pump is
installed and pump speed control circuit available, the
display “Pump=Low/(High)” may be toggled by pressing the
[N/-]
key. Note that changes made here are only for testing
the pump condition and do not become saved in the
nonvolatile memory (see Section 4.7 to update the pump
speed setting).
13) Display Battery Duration Time
This display shows the run time of the last time the unit
powered itself off due to a low battery. It is useful for
battery life testing.
14) Communicate with PC?
This display give the option of communicate with the PC for
upload or download activity.